Mini Review - Journal of Biomedical Imaging and Bioengineering (2018) Volume 2, Issue 1
Holographic microscopy for biomedical imaging.
Dario Donnarumma*
Department of Physics and Biological Systems,University of Montpellier, Montpellier, France
- *Corresponding Author:
- Dario Donnarumma
Department of Physics and Biological Systems
University of Montpellier, France
E-mail: dario.donnarumma.ita@gmail.com
Accepted date: June 06, 2018
Citation: Donnarumma D. Holographic microscopy for biomedical imaging. J Biomed Imag Bioeng. 2018;2(1):82-84.
Introduction
Blood flow imaging (BFI) is nowadays widely used in many biomedical applications, like ophthalmology [1,2] and cardiology [3], since its skill to assess physiological processes or for early detection of disease. However, many BFI techniques require the use of a contrast agent for imaging purposes, making blood flow characterization invasive [4,5]. Scanning laser-Doppler imaging can be considered to alleviate this issue by losing in acquisition speed due to the scanning process [6]. The imaging of blood flow dynamics in real time can be achieved by analyzing the spatial statistics of the dynamic speckle with the Laser Speckle Contrast Analysis/ Imaging (LASCA/LSCI) technique [7-9]. Improvement of the acquired contrast image has been achieved through exposure time optimization [10], or intensity fluctuation analysis [11] resulting in high quality perfusion images. However, these techniques are limited to perfusion monitoring in 2D and unable to reconstruct the structure of the blood flow in 3D or 4D. Recent developments in Digital Holography and Laser Doppler Holography techniques can be considered to alleviate this issue [12].
Holography invention and developement
Proposed in 1948 by Gabor [13] as an optimization of electron microscopy, classical holography aims to record, on a photosensitive material, the interference between a reference field and a diffracted object field. However, within the proposed common-path interferometric configuration, holographic imaging suffers from the so-called twin-image noise. Leith and Upatnieks [14] improve this technique by introducing an off- axis reference field, which leads to a separation of the object and its twin-image contribution in spatial frequencies domain. It is then possible to extract, from this interference pattern, the scattered field that reaches the holographic detector. This is particularly easy in digital holography, by using a camera detector. Development and democratization of high-resolution photodiode sensors played a major role in holographic imaging diffusion, making possible both digital recording and processing of holographic images [15]. Further development conduct to the introduction of heterodyne digital holography [16], in which the reference field is dynamically phase-shifted with respect to the object field. Therefore, the recorded hologram is time modulated, enabling phase-shifted interferometric measurements [17]. Temporal modulation feature of digital heterodyne holography can also be used to investigate dynamic phenomena, and being considered as a laser-Doppler imaging technique [18]. The ability of heterodyne digital holography to perform Doppler imaging has been demonstrated really valuable to investigate in vivo vasculature assessment, without contrast agent [19-21]. So, thanks to these developments in recent time, laser-Doppler holography and transmission microscopy can be coupled to investigate blood flow microcirculation by adapting a laser- Doppler holographic setup to a standard bio-microscope [22,23].
Challenges in 3D imaging
Since the Maxwell equations can be used to back propagate the field from the camera to any point of the 3D space [15], holography is an intrinsically 3D technique: the acquired hologram allows to localize and track a point-like object that scatters light, since its position coincides with the maximum of intensity of the reconstructed field [24]. The localization accuracy along the z-axis depends strongly on the experimental conditions, in particular on the detection numerical aperture (NA). Nanometric accuracy can be reached in holographic microscopy with a high numerical aperture (NA ≥ 1) objective [25,26].
By now, digital holography has been confirmed as a useful technique to image and qualitatively analyze blood flows in 2D [27,28]. Moving to the next step of 3D imaging is challenging for many of reasons. The biologically relevant size of the microcirculation makes it impossible to use a high NA objective, since the field of view would be too small. Blood cells are much larger than the wavelength, and their refractive index is close to the plasma one. The light is then scattered within a small angle in the forward direction. In transmission geometry, this corresponds to a small NA giving a poor accuracy along the z-axis. In reflection geometry, this leads to a poor signal, which competes with light scattered by the surrounding living tissues, whose refractive index is not homogeneous in time and space. These challenges has been an improvement path that is still wide open and each improvement in this direction can open other views in several applications.
Real-time 3D holography
Some examples of real time 3D holography are the in-line holographic setup for plankton in-situ observation proposed by Pfitsch et al. [29] and a parallel phase-shifting digital holography setup developed by Awatsuji et al. [30] that has been applied to the 3D reconstruction of transparent objects. Due to the complexity of using two different setup and the long acquisition time, both the above methods lack 3D reconstruction in real time.
Recently, moving red blood cells in a preclinical model have been imaged in real-time in 3D [12] with the help of a technique that combines digital holography, illumination of the sample and reconstruction along several axis [31], and calculations that involve both standard holographic propagation and 3D reconstruction by a cleaning algorithm [32-34]. More specifically, Kim et al. [35] use interferometric microscopy on mice with a method to filter out static scattering signals. The results show 3D refractive index tomograms of individual red blood cells flowing through micro capillaries. However, in this work, the part of a thin mesentery tissue should be placed on the sample holder.
Another application for zebrafish embryos is presented by Donnarumma et al. [23] and consists in a holographic microscopy working in transmission. Two illumination beams oriented in two different directions have been used to enhance the slicing along the z-axis; the angle between these two beams can variate between 30° and 90° [23,36]. Thus, each illumination beam signal has been selected with the aim to obtain two holograms from which two 3D intensity maps have been calculated. The highest values of correlation of these intensity maps have been selected by using a greedy algorithm to obtain a 3D reconstructed image. This procedure has been repeated over time to obtain a 4D reconstruction of the moving objects. In comparison with other in vivo imaging techniques dedicated to blood flow monitoring, including laser-Doppler and laser-speckle, the proposed approach has the major advantage of evaluating complex 3D aspects of the blood cells movements, specifically due to the recording of holograms. The proposed method can be still improved by using a more efficient cleaning algorithm for the 3D reconstruction and a setup allowing a larger angle of separation of the two illuminations beams [36,37]. These improvements should yield faster calculations and higher volume resolution.
Conclusion
In conclusion, multimodal techniques that combines digital holography, dual illumination of the sample and reconstruction with a greedy algorithm are rapidly developing, allowing 3D imaging of microcirculation over time with the possibility to evaluate physiologic parameters and investigate pathologies and anomalies. By now, these techniques have been validated on preclinical models and in vitro cultures. Other quantitative applications of this techniques like the evaluation of the individual blood cell motion with possible coincidence of images can be also studied from sequences of reconstructed images. In the next few years, these techniques can turn out in a label-free in vivo imaging as a robust alternative to other techniques for medical imaging. Meanwhile, the image treatment developed to achieve this result consists in an improvement of general application for other medical imaging techniques as computed tomography, magnetic resonance imaging, and ultrasound scans [38].
References
- Friedman E, Krupsky S, Lane A, et al. Ocular blood flow velocity in age-related macular degeneration. Ophthalmology. 1995;102:640-6.
- Kur J, Newman EA, Chan LT. Cellular and physiological mechanisms underlying blood flow regulation in the retina and choroid in health and disease. Prog Retin Eye Res. 2012;31:377-406.
- Pase MP, Grima NA, Stough CK, et al. Cardiovascular disease risk and cerebral blood flow velocity. Stroke. 2012;43:2803-5.
- Sakurada O, Kennedy C, Jehle J, et al. Measurement of local cerebral blood flow with iodo [14c] antipyrine. Am J Physiol. 1978; 234:H 59-66.
- Kanno I, Iida H, Miura S, et al. A system for cerebral blood flow measurement using an H215O autoradiographic method and positron emission tomography. J Cereb Blood Flow Metab. 1987;7:143-53.
- Yeh Y, Cummins H. Localized fluid flow measurements with an he-ne laser spectrometer. Appl Phys Lett. 1964;4:176-8.
- Briers JD, Webster S. Laser speckle contrast analysis (lasca): a nonscanning, full-field technique for monitoring capillary blood flow. J Biomed Opt. 1996;1:174-9.
- Dunn AK. Laser speckle contrast imaging of cerebral blood flow. Ann Biomed Eng. 2012;40:367-77.
- Briers D, Duncan DD, Hirst E, et al. Laser speckle contrast imaging: theoretical and practical limitations. J Biomed Opt. 2013;18:066018.
- Yuan S, Devor A, Boas DA, et al. Determination of optimal exposure time for imaging of blood flow changes with laser speckle contrast imaging. Appl Opt. 2005; 44:1823-30.
- Zeng Y, Wang M, Feng G, et al. Laser speckle imaging based on intensity fluctuation modulation. Opt Lett. 2013;38:1313-5.
- Donnarumma D, Brodoline A, Alexandre D, et al. 4D holographic microscopy of zebrafish larvae microcirculation. Opt Express. 2016;24:26887-900.
- Gabor D. A new microscopic principle. Nature. 1948;161:777-8.
- Leith EN, Upatnieks J. Reconstructed wavefronts and communication theory. J Opt Soc Am. 1962;52:1123-8.
- Schnars U, Juptner W. Direct recording of holograms by a CCD target and numerical reconstruction. Appl Opt. 1994;33:179-81.
- Clerc F, Collot L, Gross M. Numerical heterodyne holography with two-dimensional photodetector arrays. Opt Lett. 2000;25:716-8.
- Atlan M, Gross M, Absil E. Accurate phase-shifting digital interferometry. Opt Lett. 2007;32:1456-8.
- Atlan M, Gross M. Laser Doppler imaging, revisited. Rev Sci Instrum. 2006;77:116103.
- Atlan M, Forget BC, Boccara AC, et al. Cortical blood flow assessment with frequency-domain laser Doppler microscopy. J Biomed Opt. 2007;12: 024019.
- Atlan M, Gross M, Forget BC, et al. Frequency-domain wide-field laser Doppler in vivo imaging. Opt Lett. 2006;3:2762-4.
- Simonutti M, Paques M, Sahel JA, et al. Holographic laser Doppler ophthalmoscopy. Opt Lett. 2010;35:1941-3.
- Alexandre D, Lutfalla G, Gross M. Holographic imaging of Zebrafish embryo blood flow with dually oriented illumination beams. In Digital Holography and Three-Dimensional Imaging. Optical Society of America. 2015.
- Donnarumma D, Brodoline A, Alexandre D, et al. Blood flow imaging in zebrafish by laser doppler digital holography. Microsc Res Tech. 2018;81:153-61.
- Atlan M, Gross M, Desbiolles P, et al. Heterodyne holographic microscopy of gold particles. Opt lett. 2008;3:500-2.
- Lee SH, Roichman Y, Yi GR, et al. Characterizing and tracking single colloidal particles with video holographic microscopy. Opt Express. 2007;15:18275-82 .
- Cheong FC, Krishnatreya BJ, Grier DG. Strategies for three-dimensional particle tracking with holographic video microscopy. Opt Express. 2010;18;13563-73.
- Gao J, Lyon JA, Szeto DP, et al. In vivo imaging and quantitative analysis of zebrafish embryos by digital holographic microscopy. Biomed Opt Express. 2012;3:26230-5.
- Verrier N, Alexandre D, Gross M. Laser doppler holographic microscopy in transmission: application to fish embryo imaging. Opt Express. 2014;22:9368-79 .
- Pfitsch DW, Malkiel E, Takagi M, et al. Analysis of in-situ microscopic organism behavior in data acquired using a free-drifting submersible holographic imaging system. IEEE. 2007;1-8.
- Awatsuji Y, Wang Y, Xia P, et al. Parallel phase-shifting digital holography system using dual polarization-imaging cameras for 3D imaging of transparent dynamic object. In Laser Applications to Chemical, Security and Environmental Analysis .Optical Society of America. 2016.
- Saglimbeni F, Bianchi S, Lepore A, et al. Three-axis digital holographic microscopy for high speed volumetric imaging. Opt Express. 2014;22:13710-8.
- Hogbom J. Aperture synthesis with a non-regular distribution of interferometer baselines. Astron Astrophys Suppl Ser. 1974;15:417-26.
- Soulez F, Denis L, Fournier C, et al. Inverse-problem approach for particle digital holography: accurate location based on local optimization. J Opt Soc Am A Opt Image Sci Vis. 2007;1164-71.
- Soulez F, Denis L, Thiebaut E, et al. Inverse problem approach in particle digital holography: out-of-field particle detection made possible. J Opt Soc Am. 2007;24:3708-16.
- Kim K, Choe K, Park I, et al. Holographic intravital microscopy for 2-D and 3-D imaging intact circulating blood cells in microcapillaries of live mice. Scientific reports. 2016;6:33084.
- Donnarumma D, Rawat N, Brodoline A. High-speed quantitative 3D imaging by dual-illumination holographic microscopy. Microsc Res Tech. 2018.
- Donnarumma D, Rawat N, Brodoline A. High-speed Quantitative 3D Blood Flow Imaging by Dual-illumination Holographic Microscopy. In Digital Holography and 3-D Imaging. Optical Society of America. 2018.
- Gross M, Donnarumma D, Brodoline A. New developments in ultrasound-modulated optical tomography made by heterodyne holography. In Digital Holography and Three-Dimensional Imaging. Optical Society of America. 2016.